EasyManua.ls Logo

R&S SM300 - Functional Test

R&S SM300
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S SM300 Functional Test
Version 04 67
Service Manual SM300: 1147.1652.00 Instrument Part No SM300: 1147.1498.03
Significant values
difference error
Occurs if values from an instrument are not identical to the default values
specified for this instrument, i.e. part numbers and names.
Click <OK> to continue and run the Series 300 Wizard to enter all
modifications made to the instrument.
After you have exited the wizard and performed Read from instrument,
this window should not appear again.
User abort error
Occurs if you cancel an action (e.g.
Waiting for instrument status). The
following window will appear:
Click <OK> to continue.
1.6 Functional Test
Once the spare parts have been replaced, check that the instrument is
functioning correctly:
1. Run the self test (result must be passed).
2. Check the spinwheel and each button on the keypad.
3. Check the USB bus interfaces with the Series 300 module registration
software. If this software is running, the USB interface has been
checked.
4. Check the reference frequency as described in section 2.4.3
5. Check the RF level as described in section 2.4.4.1 "Test method for level
in measurement range of power meter"
6. Check the time and date, and correct them if necessary.
7. Check whether all messages have been deleted (see section 5.2.2).
8. Set instrument to the factory default settings (preset).

Table of Contents

Related product manuals