Module Registration Software R&S SM300
Version 04 66
Service Manual SM300: 1147.1652.00 Instrument Part No SM300: 1147.1498.03
Click <OK> to set the correct value
Wait for “PROGRAM IDLE” status
Restart the instrument
Serial number error
Occurs during “Read from Instrument” action if the measurement module or
PC module does not contain the serial number.
If the corresponding serial number for this module was found, following
window will appear:
Click <OK> to the value displayed in the message
Wait for “PROGRAM IDLE” status
Restart the instrument
If the corresponding serial number for this module was not found, following
window will appear:
Enter correct serial number
Click <OK> to set the entered value
Wait for “PROGRAM IDLE” status
Restart the instrument
Firmware release
error
Occurs if the combination of firmware versions of the MMI, measurement
module and power supply does not correspond to any of the firmware
releases listed in the configuration file. There are two possible reasons:
The configuration file is out-of-date. Get the latest version of the Series
300 module registration software or the updated .ini file (SM300-
releases.ini in directory \data) supplied with the software release. This
combination of firmware versions is not allowed. Update firmware of the
whole instrument to a regular release.