5.2.3 Device Error Messages (All Messages)
Possible error
messages
The MMI receives error messages from all installed USB devices (SM300
generator module and the power supply). The following list describes the possible
error messages sent from the modules.
5.2.3.1 SM300 Module
EventId Name Parameters Condition
0x6010 OSE Internal Error wParam … error code Operating system internal error
0x6011 Device Error lParam … error code Firmware or hardware error
0x6020 InitDone (none) Module initialization is done
0x6031 Single Sweep
Finished
wParam = 255 Single Sweep finished
0x6040 Pll Unlocked (none) Ref Generator is unlocked
0x6050 Temperature Out Of
Limits
wParam:
bit 0 = 1 … AB under limit
bit 1 = 1 … AB over limit
bit 8 = 1 … DB under limit
bit 9 = 1 … DB over limit
Temperature on the boards is out of
operational limits
0x6100 Self Test Finished (none) Required self test action is finished
0x6200 MO Level
Calibration Finished
( not implemented )
(none) Required MO Level Calibration is finished
0x6300 EEPROM Destruct
Finished
lParam = 0 … no error
lParam = 1 … bad
EEPROM id
lParam = 2 … I2C error
lParam = 3 … EEPROM
error
Destructive EEPROM test is finished
0x67FF Control Transfer
Handshake
wParam … GDI Identifier
lParam = 0 (GDI_NAK)
lParam = 1 (GDI_ACK)
GDI identifier associated with the control data
transfer
GDI_ACK is always sent with the event
0x6F00 Interrupt Data Block wParam = BlockId