Command reference
R&S
®
ZNL/ZNLE
1031User Manual 1178.5966.02 ─ 20
Tip: Use CALCulate<Chn>:MARKer<Mk>:BWIDth to set the bandwidth and query
the results of a bandfilter search. Note the sign convention for input values.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk> This numeric suffix is ignored and may be set to any value
because the bandfilter search functions always use markers M1
to M4.
Parameters:
<Bandwidth> Range: -100 dB to 100 dB
Increment: 0.03 dB
*RST: 3 dB
Default unit: dB
Example: See CALCulate<Chn>:MARKer<Mk>:BWIDth
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DELTa:STATe <Boolean>
Switches the delta mode for marker <Mk> on trace no. <Chn> on or off.
Note: This command is the ZVR-compatible equivalent of CALCulate<Chn>:
MARKer<Mk>:DELTa[:STATe].
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk> Marker number.
Parameters:
<Boolean> ON | OFF - enable or disable the delta mode.
*RST: OFF
CALCulate<Chn>:MARKer<Mk>:FUNCtion[:SELect] <Mode>
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion[:SELect] <Mode>
Selects a search mode for the related marker.
The search can then be initiated using one of the commands:
CALCulate<Ch>:MARKer<Mk>:SEARch...
CALCulate<Ch>:MARKer:REFerence:SEARch...
CALCulate<Ch>:MARKer<Mk>:MAXimum or
CALCulate<Ch>:MARKer<Mk>:MINimum
CALCulate<Ch>:MARKer:REFerence:MAXimum or
CALCulate<Ch>:MARKer:REFerence:MINimum
The marker must be created before using CALCulate<Chn>:MARKer<Mk>[:STATe]
ON or CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON.