Command reference
R&S
®
ZNL/ZNLE
1032User Manual 1178.5966.02 ─ 20
Note: This command is not needed except for compatibility with ZVR programs. Use
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute or CALCulate<Chn>:
MARKer<Mk>:REFerence:FUNCtion:EXECute to select a search mode and at the
same time initiate the search.
Suffix: 
<Chn>
.
Channel number used to identify the active trace.
<Mk> Marker number.
Ignored for reference markers and for bandfilter searches
(BFILter), because bandfilter searches always use markers
M1 to M4.
Parameters:
<Mode> MAXimum | MINimum | RPEak | LPEak | NPEak | TARGet |
LTARget | RTARget | BFILter | MMAXimum | MMINimum |
SPRogress
See CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
CALCulate<Chn>:MARKer<Mk>:FUNCtion:TARGet <SearchValue>
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:TARGet <SearchValue>
Defines the target value for a target search with the related marker (see "Marker
addressing" on page 681), which can be activated using CALCulate<Chn>:
MARKer<Mk>:FUNCtion:EXECute TARGet or CALCulate<Chn>:MARKer<Mk>:
REFerence:FUNCtion:EXECute TARGet.
Note: These commands are the ZVR-compatible equivalents of CALCulate<Chn>:
MARKer<Mk>:TARGet and CALCulate<Chn>:MARKer<Mk>:REFerence:TARGet.
Suffix: 
<Chn>
.
Channel number used to identify the active trace.
<Mk> Marker number (ignored for reference markers)
Parameters:
<SearchValue> Target search value of marker no. <Mk>.
Range:  Depending on the format of the active trace. For a
dB Mag trace the range is -200 dB to +200 dB.
Increment:  0.1 dB
*RST:  Depending on the trace format; 0 dB for a dB Mag
trace.
Default unit: dB
CALCulate<Chn>:MARKer<Mk>:MAXimum
CALCulate<Chn>:MARKer<Mk>:MINimum
CALCulate<Chn>:MARKer<Mk>:REFerence:MAXimum
CALCulate<Chn>:MARKer<Mk>:REFerence:MINimum
These commands select a search mode for the related marker (see "Marker address-
ing" on page 681) and initiate a maximum or minimum search.