Command reference
R&S
®
ZNL/ZNLE
1044User Manual 1178.5966.02 ─ 20
Calibration type Parameter in
[SENSe<Ch>:]CORRec-
tion:COLLect:METHod
Available error terms (depending on port
numbers)
TOSM TOSM 'SCORR1' to 'SCORR12'
(at present the isolation terms 'SCORR4' and
'SCORR10' are not included)
TOM, TSM, TRM, TRL,
TNA
TOM | TRM | TRL | TNA 'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK',
'LOADMATCH', 'TRANSTRACK' (for reading
and writing)
'G11' ... 'G22' and 'H11', 'H12', 'H22' (for read-
ing only; the 'H21' matrix elements are nor-
malized to 1)
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:MAGNitude <OffsetLoss>
Defines the frequency-independent part (DC value) of the offset loss.
Tip: Use the [SENSe<Ch>:]CORRection:LOSS<PhyPt>... commands to define
the complete set of loss offset parameters.
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:MAGNitude is equivalent to
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:OFFSet.
Suffix: 
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer
Parameters:
<OffsetLoss> Frequency-independent part of the offset loss
Range:  -200 dB to +200 dB
Increment:  0.0001 dB
*RST:  0 dB
Default unit: dB
[SENSe<Ch>:]FREQuency:MODE <FreqSweep>
Selects the sweep type and defines which set of commands controls the stimulus fre-
quency.
Tip: The command [SENSe<Ch>:]SWEep:TYPE provides a complete list of sweep
types.
Suffix: 
<Ch>
.
Channel number
Parameters:
<FreqSweep> CW | FIXed | SWEep | SEGMent
Linear or logarithmic frequency sweep, depending on the
selected spacing ([SENSe<Ch>:]SWEep:SPACing LINear |
LOGarithmic). The frequency range is set via [SENSe<Ch>:
]FREQuency:STARt etc.
VNA command reference