Command reference
R&S
®
ZNL/ZNLE
934User Manual 1178.5966.02 ─ 20
Example: 
*RST; :CORR:LOSS:AUTO ONCE
Reset the instrument and apply the "Auto Length and Loss"
function to the default trace (Trc1 in channel 1).
Usage:  Setting only
Manual operation:  See "Auto Length and Loss" on page 549
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:FREQuency <RefFreq>
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond:FREQuency <RefFreq>
These commands define the reference frequencies f
1
 and f
2
 for the offset loss approxi-
mation.
The corresponding losses can be defined using [SENSe<Ch>:]CORRection:
LOSS<PhyPt>:OFFSet.
Suffix: 
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<RefFreq> Reference frequency
Range:  Frequency range of the analyzer model.
Increment:  1 MHz
*RST:  1000000000 Hz (= 1 GHz)
Default unit: Hz
Manual operation:  See "Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss" on page 534
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:OFFSet <OffsetLoss>
Defines the frequency-independent part (DC value) of the offset loss.
Suffix: 
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer
Parameters:
<OffsetLoss> Frequency-independent part of the offset loss
Range:  -200 dB to +200 dB
Increment:  0.001 dB
*RST:  0 dB
Default unit: dB
Example:  See [SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation:  See "Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss" on page 534
VNA command reference