EasyManua.ls Logo

R&S ZNLE Series - Page 950

R&S ZNLE Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNL/ZNLE
950User Manual 1178.5966.02 ─ 20
[SENSe<Ch>:]SEGMent<Seg>:DEFine:SELect................................................................ 955
[SENSe<Ch>:]SEGMent<Seg>:DELete:ALL..................................................................... 956
[SENSe<Ch>:]SEGMent<Seg>:DELete[:DUMMy]............................................................. 956
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:CENTer?.......................................................957
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:SPAN?..........................................................957
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt...........................................................957
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP........................................................... 957
[SENSe<Ch>:]SEGMent<Seg>:INSert............................................................................. 958
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect..................................................................960
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel].................................................................960
[SENSe<Ch>:]SEGMent:POWer[:LEVel]:CONTrol.............................................................961
[SENSe<Ch>:]SEGMent<Seg>[:STATe]........................................................................... 961
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl.................................................................962
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol...................................................962
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts................................................................ 963
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME...................................................................963
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:CONTrol..................................................... 964
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:SUM?.........................................................965
[SENSe<Ch>:]SEGMent<Seg>:ADD
Inserts a new sweep segment using default channel settings ("Insert New Segment").
The added segment covers the frequency interval between the maximum frequency of
the existing sweep segments and the stop frequency of the entire sweep range.
Tip: Use [SENSe<Ch>:]SEGMent<Seg>:INSert to create a segment with specific
channel settings.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number. Segment numbers must be sequential.
If n segments exist already, the added segment must have the
segment number n+1.
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM2:ADD
Create a second sweep segment. The frequency range of the
second segment will be between 1.5 MHz and the maximum fre-
quency of the analyzer.
Usage: Event
Manual operation: See "Add / Insert / Delete / Delete All" on page 335
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution] <ResBandwidth>
Defines the IF bandwidth of the analyzer (Meas. Bandwidth). Values between 1 Hz and
500 kHz can be set.
VNA command reference