256823346-R3
ERVICE
S
Parameter
Number Function
Software
Version
Default Settings
Available Setting OptionsEU US
90
Default settings without changing
calibration
2.15 & before
Reset current parameter settings to
default (factory) settings.
NOTE: Calibration settings will not
change.
2.18
91
Default settings with changing
calibration
2.15 & before
Reset current parameter and calibration
settings to default (factory) settings.
Refer to Section 3.1-6
2.18
92 A
Recall stored settings calibration 2.15 & before
View (recall) current calibration settings.
Refer to Section 3.1-4.
Convert calibration settings when
Parameter 1 has changed. Refer to
Section 3.1-5.
2.18
93 A
Read out last 10 error messages 2.15 & before
Read out last 50 error messages with
timestamp
2.18
NOTE: Actual registration of the 50 er-
ror messages with timestamp the date
and time must be set first.
94 A
Back up initial parameter settings and
calibration (factory use only)
2.18 00000 00000 NOTE: This is password protected.
95 A
Factory use only 00000 00000
96 Printout parameter setup Pr-C1 Pr-C1 Pr-C1, Pr-C2
97 Key Test Function (buzzer and nr)
98 Scale ID number 1 1 0 – 9999
99 Software version 2.XX 2.XX
A Some software versions may not offer this parameter.