3.1-3 Parameter Table (continued)
Parameter
Number Function
Software
Version
Default Settings
Available Setting OptionsEU US
63 Auto shut-off time forks 3 3
0 = never off
1 = 30 min
2 = 1 hour
3 = 2 hour
64 Sleep time indicator if used 0 0 0 (not used), 1 – 99 min
65 Auto shut-off time backlight 20 20 off, 20, 40, 80, 160, 320 sec
66 Backlight brightness 100 100
100%, 75%, 50%, 25%, 0%
(NOTE: 0% means the backlight is off)
67 Low bat switch off time forks 10 10 0 = never off, 1 – 99 min
68 Buzzer function active
Before 2.15 off off off, on = direct OR on = 2 sec delay
2.15 & 2.18 no no Yes, No
69
Disable buttons
NOTE: Use this option only if P16 is set
to one of following numbers: 1 - 14.
A
0 0
0 = All buttons and additional options
are enabled
1 = Disable Tare function
2 = Disable Tare and Sum functions
3 = Disable Tare, Sum and Entry
Mode functions
70 Tare clear 2.18 no no
No = Tare remains in memory after
new weight has been picked
Yes = Tare is cleared after new
weight is picked (NOTE: This is
default when P13 is selected as
"OIML".)
80 Digital corner calibration active 0 0
0 = enabled
1 = disabled
81 Compensation factor digital corner A 1.000 1.000
Do not change value after the forks
are calibrated (corner calibration).
82 Compensation factor digital corner B 1.000 1.000
Do not change value after the forks
are calibrated (corner calibration).
83 Compensation factor digital corner C 1.000 1.000
Do not change value after the forks
are calibrated (corner calibration).
84 Compensation factor digital corner D 1.000 1.000
Do not change value after the forks
are calibrated (corner calibration).
85
Pairing iForks to the display via
bluetooth connection (Refer to Section
3.3-1)
NOTE: Only enter this parameter to
establish a complete new bluetooth
connection. Enter last 6 characters
of MAC address of Fork 1 and Fork 2
or cancel by pressing "CLR" button.
A Some software versions may not offer this parameter.