8 check sum error in the data record of the dark offset repeat dark adjustment
9 check sum error in the data record of the device basic
setting (status, ADC correction)
automatic error remedy
10 division by a too small value (< 0.001 A) check filter; check standard;
check measuring solution
11 invalid calibration curve Select valid number
12 setting to zero not possible (zero value is < 32768 cycle) check lamp; check filter;
check zero solution
13 setting to zero not possible (zero value is > 983039 cycle) check lamp; check filter;
check zero solution
14 invalid standard measure valid standard solution
15 no parameter memory vacant (too little memory for non-
linear methods)
delete a no longer actual
nonlinear
method
16 method no. is occupied select other method no.; delete a no
longer actual
nonlinear method
17 check sum error in the parameter memory (nonlinear
method)
program method newly
18 at calculation overflows in nonlinear method check factor; check parameter
19 clock malfunction
20 overflow at Kinetic check measuring solution
21 overflow at Kinetic check measuring solution
22 overflow at Kinetic check measuring solution
23 overflow at Kinetic check measuring solution
24 overflow at Kinetic check measuring solution
25 overflow at Kinetic check measuring solution
26 overflow at Kinetic check measuring solution
27 overflow at Kinetic check measuring solution
28 overflow at Kinetic check measuring solution
29 wrong input of deltas or time per delta restart method
30 battery empty contact service partner
31 communication: wrong data format contact service partner
32 communication: sent data not plausible / not interpretable contact service partner
33 communication: mentioned module does not answer in a
certain time
check interconnecting cable;
check mentioned module
34 communication: overflow send/receive buffer reduce amount of data at the
communication partner
35 remote control: wrong method number external software problem
36 remote control: unknown command external software problem
37 remote control: wrong data format external software problem
38 check sum of operating system bank 0 damaged contact service partner
39 check sum of operating system bank 1 damaged contact service partner
40 timeout at reception from module 2 switch off/on device
41 check sum error at reception from module 2 switch off/on device
42 NAK at reception from module 2 switch off/on device
ROBERT RIELE GmbH & Co KG Page 65 9 - ERROR MESSAGE / CORRECTION