5010e_513.doc / 22.01.10
44 error at pump calibration
- pump tube not clamped;
- no liquid aspirated;
- too much liquid aspirated during calibrating;
- power of the pump too small;
- motor does not rotate or stops;
- connectors of tubes leaky;
- bubble detector works falsely;
- aspiration tube blocked;
- aspiration tube badly stained
- clamp pump tube;
- control volume of aspiration;
- control connectors of tubes;
- the plug of the bubble detector has a bad
contact;
- the cable of the bubble detector has a
bad contact;
- adjust bubble detector (chapter 7.2.4);
- replace aspiration tube (chapter 8.7)
45 error at sipping solution
- pump tube not clamped;
- no liquid aspirated;
- power of the pump too small;
- motor does not rotate or stops;
- connectors of tubes leaky;
- bubble detector works falsely;
- aspiration tube blocked;
- aspiration tube badly stained
- clamp pump tube;
- adjust bubble detector (chapter 7.2.4);
- control volume of aspiration;
- control connectors of tubes;
- do a calibration of peristaltic pump
(chapter 7.2.4);
- the plug of the bubble detector has a bad
contact;
- the cable of the bubble detector has a
bad contact;
- replace aspiration tube (chapter 8.7)
46 filter position out of tolerance
47 + 12 V
DC
out of range contact service partner
51 Wrong adaptor Insert standard cuvette adaptor
(chapter3.5.1)
52 timeout printer internal the internal printer is temporarily discon-
nected
53 Set of data points is missing Check multi-standard functions
54 nonlinear measurement: value out of range check the absorbance of the sample
55 number of given data points < 2 add data points
57 memory of data logging management is completely filled delete log data (chapter 7.2.8)
58 checksum of data logging damaged delete log data (chapter 7.2.8)
59 error in the automatic measuring operation (remote) check interface
60 error at multiplexer of operational amplifier contact service partner
61 error at multiplexer of bubble detector contact service partner
62 free method number not found check method memory
63 wrong address at multi-standard loading check multi-standards
64 current method not found in monthly memory check QC data of method
65 more than 50 QC methods defined delete unused QC methods
66 internal clock is off. QC data not storable switch on internal clock
67 BCC error in dataset of QC method values check current method
68 free monthly memory not found at QC delete unused QC methods
69 free space in daily memory not found at QC empty QC daily memory
70 error at QC calculation check QC data
71 QC serum not found check QC data
72 wrong address at saving test results send results through serial interface and
delete results (chapter 7.2.19)
73 results memory full, it will be overwritten overwrite results memory or send and
delete results (chapter 7.2.19)
74 BCC error at sending result string send results through serial interface and
delete results (chapter 7.2.19)
ROBERT RIELE GmbH & Co KG Page 66 9 - ERROR MESSAGE / CORRECTION