DG1000 Command System RIGOL
Programming Guide for DG1000
2-41
Detailed information of each command:
1.
FSK:SOURce
Syntax FSK:SOURce {INTernal|EXTernal}
Function Select internal or external modulation source, the default is INT.
Example FSK:SOUR EXT
2.
FSK:SOURce?
Syntax
FSK:SOURce?
Function
Query the modulation source of FSK.
Return Value The query returns INT or EXT.
3.
FSK:FREQuency
Syntax FSK:FREQuency {<frequency>|MINimum|MAXimum}
Function Set the hop frequency of FSK in Hz.
Example FSK:FREQ 10
4.
FSK:FREQuency?
Syntax
FSK:FREQuency?
Function
Query the hop frequency of FSK.
Return Value The query returns the hop frequency of FSK in scientific notation
and in Hz, such as: 1.000000e+01.
5. FSK:INTernal:RATE
Syntax
FSK:INTernal:RATE {<rate>|MINimum|MAXimum}
Function Set
the rate at which the output frequency “shifts” between the
carrier and hop frequencies, the unit is Hz.
Explanation Rate range: 2mHz to 50kHz
Example
FSK:INT:RATE 100
6. FSK:INTernal:RATE?
Syntax
FSK:INTernal:RATE?
Function Query the FSK rate.
Return Value The query returns the FSK ratein scientific notation, such as:
1.000000e+02.
7.
FSK:STATe