RIGOL Chapter 5 Specifications
Users Guide for DP1308A
5-2
Transient Response Time
Less than 50μs is spent on recovering the voltage within 15mV during the output
current changes from full load to half load or half to full.
Command Processing Time
[2]
<50ms
Temperature Coefficient per (output percentage + offset)
Voltage
0.01%+2mV
0.01%+3mV
Current
0.02%+3mA
0.01%+0.5mA
Stability
[3]
± (output percentage + offset)
Voltage
0.03%+1mV
0.02%+2mV
Current
0.1%+3mA
0.05%+1mA
Voltage Programming Speed(to within 1% of total variation range)
Rising
Full Load
11ms
50ms
No Load
10ms
45ms
Falling
Full Load
13ms
20ms
No Load
200ms
400ms
OVP/OCP
Accuracy ± (output
percentage + offset)
0.5%+0.5V/0.5%+0.5A
Activation Time
1.5ms (OVP≥ 3V)
<10ms (OVP<3V and OCP)
Machine
Dimension
235 mm (W) x 155 mm (H) x 384 mm (D)
Weight
8.5 kg
Power Supply
AC Output
(50Hz-60Hz)
100Vac+10%, 115Vac+10%, 220Vac+10%,
230Vac+10% (250VAC Max)
Environment
Working Temperature
Full rated value output: 0℃~40
At higher temperature: the output current falls into
50% at the maximum temperature 55
Cooling Method
Fan cooling
Remarks:
[1] Specifications are for one hour warm-up and at 25.
[2] The maximum time required for regulating corresponding output when received APPLy and
SOURce commands.
[3] The variation of output within 8 hours after warm-up 30 minutes and both the load circuit and