Chapter 5 To Trigger the Oscilloscope RIGOL
DS8000-R User Guide 5-29
Data Type:
Click Data Type continuously to set the effective pattern of the data signal. It can
be set to H (high level) or L (low level).
Trigger Condition:
Click When to select the trigger condition of Setup/Hold trigger.
Setup: the oscilloscope triggers when the setup time is smaller than the
specified setup time. After selecting this type, click More Setup to set the
setup time. You can use the pop-up numeric keypad or scroll with the mouse to
set it.
Hold: the oscilloscope triggers when the hold time is smaller than the specified
hold time. After selecting this type, click More Hold to set the hold time. You
can use the pop-up numeric keypad or scroll with the mouse to set it.
Setup/Hold: the oscilloscope triggers when the setup time or the hold time is
smaller than the specified time value. After selecting this type, click More
Setup and More Hold to set the setup time and hold time respectively. You
can use the pop-up numeric keypad or scroll with the mouse to set them.
Trigger Mode:
For details, refer to descriptions in "Trigger Mode".
Trigger Parameter Setting:
Set the trigger parameter (noise rejection) under this trigger type. For details, refer
to descriptions in "Noise Rejection".
Trigger Level:
Refer to "Trigger Level". The current trigger level value is displayed at the
upper-right corner of the screen.