Chapter 6 Specifications
DC Characteristics
Accuracy Specifications: ± (% of reading + % of range)
[1]
Function Range
[2]
Test Current or
Load Voltage
24 Hour
[3]
T
CAL
±1
90 Day
T
CAL
±5
1 Year
T
CAL
±5
0
to (T
CAL
-5
)
DC Voltage
200.0000V
0.0020 + 0.0006 0.0040 + 0.0006 0.0050 + 0.0006 0.0005 + 0.0001
300.000V
0.0020 + 0.0006 0.0040 + 0.0010 0.0055 + 0.0010 0.0005 + 0.0001
DC Current
Resistance
[4]
Specifications are for 90-minute warm-up and 100 PLC integration time.
10% overrange on all ranges.
Relative to calibration standards.
Specifications are for 4–wire resistance measurement. Add 3 Ω additional error in 2-wire resistance measurement.