Chapter 6 Specifications RIGOL
M300 User’s Guide
6-1
Chapter 6 Specifications
DC Characteristics
Accuracy Specifications: ± (% of reading + % of range)
[1]
Function Range
[2]
Test Current or
Load Voltage
24 Hour
[3]
T
CAL
±1
90 Day
T
CAL
±5
1 Year
T
CAL
±5
Temperature Coefficient
0
to (T
CAL
-5
)
(T
CAL
+5) to 50
DC Voltage
200.0000mV
-
0.0020+ 0.0020
0.0030 + 0.0025
0.0040 + 0.0025
0.0005 + 0.0005
2.000000V
-
0.0015 + 0.0005
0.0020 + 0.0006
0.0035 + 0.0006
0.0005 + 0.0001
20.00000V
-
0.0020 + 0.0004
0.0030 + 0.0005
0.0040 + 0.0005
0.0005 + 0.0001
200.0000V
-
0.0020 + 0.0006 0.0040 + 0.0006 0.0050 + 0.0006 0.0005 + 0.0001
300.000V
-
0.0020 + 0.0006 0.0040 + 0.0010 0.0055 + 0.0010 0.0005 + 0.0001
DC Current
200.0000µA
<0.03V
0.010 + 0.012
0.040 + 0.015
0.050 + 0.015
0.0020 + 0.0030
2.000000mA
<0.25V
0.007 + 0.003
0.030 + 0.003
0.050 + 0.003
0.0020 + 0.0005
20.00000mA
<0.07V
0.007 + 0.012
0.030 + 0.015
0.050 + 0.015
0.0020 + 0.0020
200.0000mA
<0.7V
0.010 + 0.002
0.030 + 0.003
0.050 + 0.003
0.0020 + 0.0005
1.000000A
<0.12V
0.050 + 0.020
0.080 + 0.020
0.100 + 0.020
0.0050 + 0.0010
Resistance
[4]
200.0000Ω
1mA
0.0030 + 0.0030
0.008 + 0.004
0.010 + 0.004
0.0006 + 0.0005
2.000000kΩ
1mA
0.0020 + 0.0005
0.008 + 0.001
0.010 + 0.001
0.0006 + 0.0001
20.00000kΩ
100µA
0.0020 + 0.0005
0.008 + 0.001
0.010 + 0.001
0.0006 + 0.0001
200.0000kΩ
10µA
0.0020 + 0.0005
0.008 + 0.001
0.010 + 0.001
0.0006 + 0.0001
1.000000MΩ
2µA
0.002 + 0.001
0.010 + 0.001
0.012 + 0.001
0.0010 + 0.0002
10.00000MΩ
200nA
0.015 + 0.001
0.030 + 0.001
0.040 + 0.001
0.0030 + 0.0004
100.0000MΩ
200nA || 10MΩ
0.300 + 0.010
0.800 + 0.010
0.800 + 0.010
0.1500 + 0.0002
[1]
Specifications are for 90-minute warm-up and 100 PLC integration time.
[2]
10% overrange on all ranges.
[3]
Relative to calibration standards.
[4]
Specifications are for 4wire resistance measurement. Add 3 Ω additional error in 2-wire resistance measurement.