RIGOL Chapter 2 Front Panel Operations
2-54 M300 User’s Guide
To Control MC3164
MC3164 is a 64-channel single-ended multiplexer (MUX-64 for short). MC3164 is
divided into two banks of 32 single-ended channels (HI only, the 32 channels has a
common LO terminal) each and cannot be used for 4-wire connection. MC3164 is
used with the external terminal block (M3TB64) which has a built-in thermocouple
reference junction.
01
...
...
DMM
Source
H
L
32
33
64
Bank Switch
Channel Switches
96
Backplane Switches
Com H
Com L
95
Signals
to be
Measured
Reference
Junction
Terminal Block
: M
3
TB64
H
01
H
64
.
.
.
78
-
Pin
Interface
Com L
Figure 2-20 MC3164 Schematic Diagram
The control interface of MC3164 is as shown in Figure 2-22
. Press Page Up and
Page Dn at the front panel to open the previous and next pages of the control
interface. Use the up/down and left/right direction keys to select the desired channel.
The channel currently selected is highlighted, for example,
. You can perform
various operations on the channel currently selected (such as disconnect and close
the channel).