Chapter 2 Front Panel Operations RIGOL
M300 User’s Guide
2-53
5 Bank Separate/Bank Joined
By default, this module is in “Bank Joined” state (namely the bank switch in
Figure 2-17 is closed and the control interface is as shown in Figure 2-18
). At
this point, this module can only be used for 2-wire connection. Press Page Dn
Bank Sepa to execute the Bank Separateoperation (namely disconnect
the bank switch in
Figure 2-17 and the control interface is as shown in Figure
2-19). At this point, this module is divided into two independent banks and can
be used for 4-wire connection. When it is used for 4-wire resistance
measurement, the instrument pairs channel n and channel n+16 automatically
(wherein, n is an integer between 1 and 16). Channel n is used to connect the
source terminal of DMM and channel n+16 is used to connect the sense terminal
of DMM.
Figure 2-19 MC3132 Control Interface (Bank Separate)