RIGOL Chapter 6 Specifications
6-2 M300 User’s Guide
Measuring Characteristics
200mV, 2V, 20V ranges: 10MΩ or >10GΩ
(For these ranges, input beyond ±26 V are clamped through 106 kΩ)
200V and 300V ranges: 10MΩ±1%
℃
CMRR (common mode
rejection ratio)
140 dB for 1 kΩ unbalanced resistance in LO lead, ±300 VDC peak maximum.
4-wire or 2-wire resistance
Current source referenced to LO input
Max. Lead Resistance
(4-wire)
10% of range per lead for 200 Ω and 2 kΩ ranges, 1 kΩ per lead on all other ranges
Available on 200 Ω, 2 kΩ and 20 kΩ ranges.
Auto Zero OFF Operation (typical value)
Following instrument warm-up at the environment temperature ±1
℃
and <5 minutes, add 0.0001 % range + 2 uV error for DCV and 2 mΩ error for
resistance.
Reading settling times are affected by source impedance, cable dielectric characteristics and input signal changes. The default measurement delay can
ensure the correctness of the first reading for most measurements.
Measurement Considerations
Teflon or other high-impedance, low-dielectric absorption wire insulation is recommended for these measurements.