Chapter 6 Specifications RIGOL
M300 User’s Guide
50 kHz to 100 kHz, add 0.13% of range.
When the frequency is lower than 100 Hz, the specification of slow filter is only for sine wave input.
Specifications are for sine wave input >5% of range. For inputs from 1% to 5% of range, add 0.1% of range additional error. Specifications are
typical values for 200 uA, 2 mA, 2 A and 10 A ranges when frequency is >1 kHz.
Measuring Characteristics
AC-coupled True-RMS -- measure the ac component of input with up to 300 V DC bias on any range.
1 MΩ ± 2%, in parallel with <150 pF capacitance on any range
CMRR (common mode
rejection ratio)
70 dB, for the 1 kΩ unbalance in LO lead, <60 Hz common mode signal frequency, ±300 Vpeak maximum.
Direct coupled to the fuse and shunt; AC-coupled True RMS measurement (measure the AC component).
DC + AC current peak value <300% of range. Current with DC current component <1 Arms.
Settling Time Considerations
The default measurement delay of the multimeter can ensure the correctness of the first readings of most of the measurements. Make sure the RC
circuit of input terminal has been fully settled (about 1 s) before accurate measurement.