Reference Manual V1.19    Software Versions 4.xx 
Page 26  003R-682-119 
the factory. Each time a critical step is altered, the counter(s) will increase by one. Whenever the 
instrument is powered up, or setup mode is entered/exited, the current value in the counter(s) is 
displayed briefly (e.g. C00010).  
The Calibration Counter 
increments when trade critical 
settings in the Calibration (CAL) 
menu, marked with , are 
changed. An example of the 
counter is C.00010. 
 
The Configuration Counter 
increments when other trade 
critical settings (i.e. not in the 
CAL menu), marked with , are 
changed. An example of the 
counter is F.00009. 
The value(s) of the counter(s) is written on the tamperproof trade label on the front or top of the 
indicator for trade-certified applications and functions as an electronic seal. If any legal for trade 
settings are changed on the instrument, the current value of the calibration counter(s) will be 
different from the recorded value and the seal is broken. In this manual, items marked with  
indicate that the setting is legal for trade critical settings.  
 
6.6.  Passcodes 
The instrument has two levels of passcodes to provide a security lock on accessing Setup via 
the keypad.  
•  Full Setup Passcode  
•  Safe Setup Passcode  
The Full Setup passcode can also be used to access Safe Setup. 
6.6.1.  Full Setup Passcode  
Setting a passcode for Full Setup restricts any access to Full Setup. Refer to FULL.PC  
(Full Security Passcode for Digital Setup) page 36. 
6.6.2.  Safe Setup Passcode 
Setting a passcode for Safe Setup restricts access to Safe Setup functions. Refer to 
SAFE.PC 
(Safe Security Passcode for Digital Setup) page 36. 
6.6.3.  Setup Lock-Out 
If an attempt is made to enter Full or Safe Setup using an incorrect passcode, the instrument 
will respond with the message ENTRY DENIED and then the user will be returned to normal 
operating mode. A passcode counter has been set so that only three failed attempts can be 
made to access Full/Safe Setup. On the fourth attempt the user will be 'locked out' of 
Full/Safe setup. Should this occur the ENTER PASS prompt will not display, but instead the 
ENTRY DENIED message displays and returns the user to the normal operating mode. To 
rectify this issue the instrument must be turned off. When the instrument is turned back on the 
passcode counter is reset to zero (allowing the user to enter the correct passcode).