8.  Setup 
The instrument digital setup facilities provide the means to configure and calibrate the instrument. 
8.1.  Accessing Setup 
There are two methods to access the Setup area. For further details of menu items available 
in each setup mode, refer to the Setup Menu Quick Reference page 68. 
8.1.1.  Access using Full Setup 
The Full Setup method provides access to all functions in Setup, including legal for trade and 
calibration sensitive settings. Changes in Full Setup mode may result in the calibration 
counter being incremented. Items marked with  indicate that the setting is trade critical. 
Changes to passcodes and restoring the factory default settings can only be accessed in Full 
Setup mode. These items will however not increment the calibration counter. If an attempt is 
made to enter Full Setup using the incorrect passcode, the instrument will respond with the 
message ENTRY DENIED. Refer to Passcodes page 26 for more information. 
 
1.  Instrument on.  
2.  Press and hold both the <POWER> and <FUNCTION> keys together for two seconds.  
3.  Full Setup can also be entered by pressing a push-button accessed from the rear of 
the instrument. Refer to OPTIONS:R.ENTRY page 35 for more details. 
WARNING 
All items in all menus will be enabled in Full Setup. 
Care should be taken to avoid inadvertently altering the Build or Calibration 
settings. 
8.1.2.  Access using Safe Setup 
The Safe Setup method restricts access to the Trade Critical settings. Changes made in this 
mode will not increment the calibration counter. In this manual, items marked with  indicate 
that the setting is trade critical. If an attempt is made to enter Safe Setup using the incorrect 
passcode, or if an attempt is made to alter a trade critical setting while in Safe Setup, the 
instrument will respond with the message ENTRY DENIED. Refer to Passcodes page 26 for 
more information.