Reference Manual V1.19    Software Versions 4.xx 
Page 62  003R-682-119 
 
15.2.11. NET.HLD 
The <NET.HOLD> key is used to hold the current net weight value when the function key is 
pressed. It will adjust the tare value when the key is pressed again to keep the net weight at 
the same value as when it was held. 
15.2.12. MED.HLD 
The <MED.HOLD> key is used to hold the current gross and net weight values when the 
function key is pressed. It will adjust the zero value when the key is pressed again to keep 
the gross and net weight values the same as when they were held. 
 
15.2.13. MVV 
The <MVV> key is used to toggle between displaying the load cell mV/V value and the 
weight. 
 
 
Press and hold the < SET.PT > key to edit the target threshold associated with output1. 
The “SetPt.1” annunciator appears briefly on the display before editing mode is entered 
to indicate that the output 1 threshold is being edited. 
Press the <OK> key to enter the output 1 threshold. Change the threshold to the required 
value using the <SEL> and <EDT> keys. 
Press the <OK> key to store the new output 1 threshold. When the process is complete 
the display will move on to editing the output2 threshold. 
The “SetPt.2” annunciator appears briefly on the display before editing mode is entered to 
indicate that the output 2 threshold is being edited. 
Press the <OK> key to enter the output 2 threshold. Change the threshold to the required 
value using the <SEL> and <EDT> keys. 
Press the <OK> key to store the new output 2 threshold. When the process is complete 
the indicator will revert to the condition it held prior to the initial pressing of the < SET.PT > 
key.