Carries the information for the message. Some messages require no 
DATA (e.g. Read Commands) so the field is optional.  
Termination 
Message termination is possible in two ways.  
 : For normal communications that do not involve checksums use either a CRLF (ASCII 13, ASCII 
10) as a terminator or a semicolon (‘;’ ASCII ). There is no start-of-message delimiter: 
<Message>   
 
Ring Network Enhancement 
Instruments with software revision V2.31+ can be configured in a Ring Network via a M42xx module 
(software revision 1.01+). This requires the central computer to send additional framing characters, 
‘Echo-On’ (=<DC2> =ASCII 12 H) and ‘Echo-Off’ (=<DC4> =ASCII 14 H) around each command. 
Below is an example Ring Network command and response with two indicators: 
COMMAND 
<DC2>20110150:<CR><LF> 
<DC4> 
 
RESPONSE 
<DC2>20110150:<CR><LF> 
81110150:07/01/2030 17-29<CR><LF> 
82110150:07/01/2030 17-30<CR><LF> 
<DC4>