64
Recall Mode
Time waveform re-analysis function
Time waveform data stored with the transient store function can be re-analyzed.
To use this function, store data in analyzer mode using transient store. When
recalling these data, the data in a specified address range can be analyzed
again. The address range is specified in units of 1 address (256 data). Re-
analyzing can start from any specified data position in the specified address.
In vibration meter mode, re-analysis is not available.
1. Carry out transient store.
(Use a suitable frequency span and input range.)
2. Activate recall mode. The re-analysis function is automatically activated
at this time. The screen shows the time waveform display if the stored
data are time waveform data, and the spectrum display if the stored
data are spectrum data.
3. Select the part you want to analyze.
Set the display to time waveform.
When the address number is changed, the display position changes in
256-point intervals.
When the start position is changed, the display position changes in 1-
point intervals.
When the analysis mode is set to spectrum, re-analysis is carried out
automatically.