91xx pH/Electrolyte Analyzers 10 Service functions
Service functions
EXW WXE
9110 only!
This code extends the wash cycle following a serum or blood
measurement. Normally 10 Standard A segments are used for the
wash cycle; with EXW set, 20 Standard A segments will be used.
NTD DTN
9110 only!
When the code TMP is set, NTD prevents the temperature from
(9110 only) being reset after each sample and causes the last
temperature value to remain until changed.
TMP PMT
9110 only!
When this code is set, the [CHANGE PATIENT TEMPERATURE?]
menu will appear in the sampling process. The analyzer will perform
a temperature correction calculation to convert the measured results
for the desired patient temperature. The temperature will be reset
after each sample to 37°C unless NTD is entered. Temperature
corrections can only be performed in the range of 15 - 45°C.
Code Description 9110 9120 9130 9140 9180 9181
ALL
Enables Ca
2+
during automat. blood/
serum measuement
X
CHP
Display results in concentratio of cH
+
ions X
CDC
Clears all Service Codes XXXXXX
DEC
Additional display resolution XXXXXX
ECO
Economy mode XXXXXX
EXW
Extended wash following serum/blood
samples
X
FIF
Cal. voltages on display X X X X X
LEM
Last 20 error messges printed X X X
MGL
Display mg/dL X X X
NOB
No beep X X X
NTD
Prevents temperature from being reset
following each sample
X
QCC
QC values direct ISE measurement XXXXXX
SFC
SnapPak counter input X X
SLC
Set "Leasing counter" X X
SSC
Small sample cups X
TMP
Allows temperature correction for
measured results
X
Ta bl e B -2 Instrument / service code matrix
Set Reset Service code description
Ta bl e B -1 Service code descriptions