EasyManua.ls Logo

Roche 91 Series - Error Messages and Troubleshooting; Standard a Not Detected; Status-Leds; Status: Not Calibrated

Roche 91 Series
192 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics April 2007
Service manual · Version 2.0 D-5
91xx pH/Electrolyte Analyzers 13 Troubleshooting
Error messages and troubleshooting
Error messages and troubleshooting
Status-LEDs
The 5 status LEDs at the bottom of the chassis back panel are lit simultaneously
during normal operation. See below if any LED is blank.
Possible causes and remedies
o One ore more supply voltages are out of range or missing. Check the voltage test
points.
e
For details, see Power circuits on page A-33.
o If the voltages are correct:
O Check subsequent cicuits and cabling for obvious damage and replace, if
necessary.
O Check and replace the corresponding fuse.
o Replace the SBC board.
o Replace the power supply.
STATUS: NOT CALIBRATED
This message is only displayed when a calibration has been interrupted by pressing
NO or opening the sample door. Perform a calibration to return to the [READY]
Mode.
STANDARD A NOT DETECTED
This message is displayed when sample contact 1 (9110 only) or the sample sensor is
unable to properly detect the Standard A solution in a programmed time period. As
the analyzer draws a sample of Standard A, the leading edge of the solution is detected
by sample contact 1 (by bridging solution between both parts of sample contact 1) or
the sample sensor. The sample sensor must also detect a continuous flow of
Standard A Solution into the sample chamber, lasting several seconds, without
encountering air bubbles or air. After several seconds, the trailing edge of the
Figure D-1 Status LEDs (back panel, bottom)

Table of Contents

Related product manuals