Roche Diagnostics April 2007
A-20 Service manual · Version 2.0
3 Interface specifications 91xx pH/Electrolyte Analyzers
Interface specifications
Example data string information
Automatic Calibration Report
<sx>* AVL 9130*<cr><crlf>ELECTROLYTE ISE<cr><crlf> 03JAN92
10:51crcrlf*CALIBR REPORT*<cr><crlf><cr><crlf>
DailyMaintenance<cr><crlf>Performed Last: <cr><crlf> 02JAN92 10:35
<cr><crlf><cr><crlf> Standard A<cr><crlf>Na = -112mV (3)<cr><crlf>
K = -1392mV (3)<cr><crlf>Cl = -106mV (3)<cr><crlf><cr><crlf>
Difference A-B<cr><crlf>Na = 1402mV ( )<cr><crlf>K = 1032mV ( )
<cr><crlf>Cl = -1006mV ( ) <cr><crlf><cr><crlf>Fluid Pack:<cr><crlf>
68% Remaining<cr><crlf><cr><crlf><ex>
Serum sample report
<sx>* AVL 9130*<cr><crlf>ELECTROLYTE ISE<cr><crlf> 03JAN92
10:59<cr><crlf><cr><crlf> Name: ..........<cr><crlf> ..........
<cr><crlf>Sample: SERUM<cr><crlf><cr><crlf> Sample No.13
<cr><crlf><cr><crlf>Na= 159soh mmol/L<cr><crlf>K = 5.4 mmol/L
<cr><crlf>Cl= 122soh mmol/L<cr><crlf><cr><crlf>*PERFORM DAILY *
<cr><crlf>*MAINTENANCE ! *<cr><crlf><cr><crlf><ex>
Data Link Information (9180/9181 only)
The data link with the COMPACT 2/3 blood gas analyzer allows to combine ISE
results with pH/blood gas results on one printout. If Ca
2+
is activated on the 9180/
9181, a pH-corrected Ca
2+
value will be calculated and printed on the combined
sample report.
For connection of the 9180/9181 analyzer to the COMPACT 2/3 analyzer, the optional
Interface Kit is required. To install the kit, first turn both instruments off. Connect the
interface filter provided in the kit to the RS232 port on the 9180/9181. Then connect
one end of the cable to the interface filter, the other end to the COM 2 port on the
COMPACT 2/3. On the COMPACT 2/3, select 9180 under the COM 2 interface
options.
e
See COMPACT 2/3 Operator's Manual for details.