EasyManua.ls Logo

Roche 91 Series - Approvals; Classifications; Product Data

Roche 91 Series
192 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics April 2007
Service manual · Version 2.0 A-15
91xx pH/Electrolyte Analyzers 2 Specifications
Specifications
Humidity maximum 85 % relative humidity, non-condensing
Data management Quality control memory storage, 3 levels, 35 values; calculation of mean, standard
deviation and coefficien of variation (CV)
Diganostic programs user-controlled diagnostics, YES/NO operation via the display
Electronics Microprocessor-controlled
Display LCD dot-matrix, 2 lines, 16 characters per line
Printer Integral thermal printer, 16 characters width
Autosampler (9181 only) Integral turntable, 18 positions, 2 mL or 0.25 mL sample cups
Interface RS232C serial port
DataLink Data link to COMAPCT 2/3 (9180/9181)
Electrical requirements 100 - 240 VAC, 50/60 Hz (90 - 250 VAC for 9110), self adjusting
Nominal power consumption 50 W
Dimensions
Classifications
Product data
Safety category I
Device type B (according to ÖVE-MG/EN 60601-1,
IEC 601-1)
Mode of operation continuous operation
Protection classification IP 20
Explosion protection the device is not designed for operation in explosive environments.
Approvals
CSA, CE, FCC Class B
CLIA Complexity Category: Moderate
Test System Code: 04739
Height 335 mm (12.2")
Width 315 mm (12.4")
Depth 295 mm (12.0")
Weight approx. 6 kg (13 lbs)
Ta bl e A -4 Instrument dimensions