of 61 Host Interface Manual for use in the US
Pos No Field Type Max EV CV
Comments
Elecsys Type Format
cobas Type Format
11 (9) Operator Identification ST 6 X Indicates operator ID who conducted measurement. HOST can't
set this field.
12 (10) Date/Time Test
Started
TS 14 Indicates time and date when
starting measurement. Setting
is as follows:
YYYYMMDDHHMMSS
Indicates time and date when
completing measurement.
Setting is as follows:
YYYYMMDDHHMMSS
14 Instrument
Identification
X E1
Table 18: Test order record
cobas type format transmits calculated tests. Elecsys type format does not transmit
calculated tests.
Elecsys type format has no Review by exception function.
type format does not transmit tests specified by exception alarm, defined in
the Review by exception screen.
Communication of batch results with review by exception alarm flags is based on
option with Review by Exception Results on Send to Host window shown
below.
• On: tests specified by Exception Alarm are communicated.
• Off: tests specified by Exception Alarm are not communicated.
Figure 11: Data selection
When Not Send Host is selected on the window Send to Host, only samples which
were not communicated to the host, are transmitted.