Remote Commands
R&S
®
FSW
1023User Manual 1173.9411.02 ─ 19
Example:
CAL:RES?
would return, e.g.
Total Calibration Status:
PASSED, Date (dd/mm/yyyy): 12/07/2004,
Time: 16:24:54,Runtime: 00.06
Usage: Query only
Manual operation: See "Alignment Results" on page 577
DIAGnostic:SERVice:INPut:MC:DISTance <Bandwidth>
This command selects the distance of the peaks of the microwave calibration signal for
calibration of the YIG filter.
This command is only available for instrument models R&S FSW13/26.
Parameters:
<Bandwidth> SMALl
Small offset of combline frequencies.
WIDE
Wide offset of combline frequencies.
Manual operation: See "Calibration Frequency MW" on page 587
DIAGnostic:SERVice:INPut:PULSed:CFRequency <Frequency>
This command defines the frequency of the calibration signal.
Before you can use the command, you have to feed in a calibration signal with
DIAGnostic:SERVice:INPut[:SELect].
Parameters:
<Frequency> Possible frequencies of the calibration signal are fix.
If you define a frequency that is not available, the R&S FSW
uses the next available frequency. Example: a frequency of
20 MHz is rounded up to the next available frequency
(21.33 MHz).
*RST: 64 MHz
Default unit: Hz
Manual operation: See "Calibration Frequency RF" on page 587
DIAGnostic:SERVice:INPut:RF:SPECtrum <Bandwidth>
This command selects the bandwidth of the calibration signal.
Parameters:
<Bandwidth> NARRowband
Narrowband signal for power calibration of the frontend.
BROadband
Broadband signal for calibration of the IF filter.
Configuring the R&S FSW