General Instrument Setup
R&S
®
FSW
590User Manual 1173.9411.02 ─ 19
Relay Cycle Counter................................................................................................... 590
Temperatures..............................................................................................................590
Ventilation................................................................................................................... 590
Relay Cycle Counter
The hardware relays built into the R&S FSW may fail after a large number of switching
cycles (see data sheet). The counter indicates how many switching cycles the individ-
ual relays have performed since they were installed.
Remote command:
DIAGnostic:INFO:CCOunter? on page 1038
Temperatures
Some hardware parts fail at high temperatures. Several temperature sensors in the
R&S FSW provide the current temperature for the CPUs and the motherboard, which
are indicated here.
Ventilation
High temperatures in the R&S FSW may occur when the fans fail. The current speed
and temperatures of the built-in fans are displayed. High temperatures or very slow fan
speed may indicate a hardware problem.
9.4 Display Settings
Some general display settings are available regardless of the current application or
operating mode. For information on optimizing your display for measurement results
see chapter 7.1, "Result Display Configuration", on page 440.
The general display settings are defined in the "Display" dialog box which is displayed
when you press the SETUP key and then select "Display".