Applications and Operating Modes
R&S
®
FSW
108User Manual 1173.9411.02 ─ 19
Sequencer State
Activates or deactivates the Sequencer. If activated, sequential operation according to
the selected Sequencer mode is started immediately.
Remote command:
SYSTem:SEQuencer on page 689
INITiate:SEQuencer:IMMediate on page 688
INITiate:SEQuencer:ABORt on page 688
Sequencer Mode
Defines how often which measurements are performed. The currently selected mode
softkey is highlighted blue. During an active Sequencer process, the selected mode
softkey is highlighted orange.
"Single Sequence"
Each measurement is performed once, until all measurements in all
active channels have been performed.
"Continuous Sequence"
The measurements in each active channel are performed one after
the other, repeatedly, in the same order, until sequential operation is
stopped.
This is the default Sequencer mode.
"Channel-defined Sequence"
First, a single sequence is performed. Then, only channels in continu-
ous sweep mode are repeated.
Channel-defined sequence is not available in Multi-Standard Real-
time (MSRT) operating mode.
Remote command:
INITiate:SEQuencer:MODE on page 688
4.5.3 How to Set Up the Sequencer
In order to perform the configured measurements consecutively, a Sequencer function
is provided.
1. Configure a channel for each measurement configuration as required, including the
sweep mode.
2. In the toolbar, select the "Sequencer" icon.
The "Sequencer" menu is displayed.
3. Toggle the "Sequencer" softkey to "On".
A continuous sequence is started immediately.
Running a Sequence of Measurements