CALC:MARK:FUNC:HARM:DIST? TOT
//Queries the total harmonic distortion.
11.5.11 Measuring the Third Order Intercept Point
● Determining the TOI..............................................................................................799
● Programming Example: Measuring the TOI..........................................................800
11.5.11.1 Determining the TOI
All remote control commands specific to TOI measurements are described here.
Useful commands for TOI measurements described elsewhere
● CALCulate<n>:DELTamarker<m>:X on page 931
● CALCulate<n>:DELTamarker<m>:X:RELative? on page 942
● CALCulate<n>:DELTamarker<m>:Y? on page 943
● CALCulate<n>:MARKer<m>:X on page 932
● CALCulate<n>:MARKer<m>:Y? on page 943
Remote commands exclusive to TOI measurements
CALCulate<n>:MARKer<m>:FUNCtion:TOI[:STATe]......................................................... 799
CALCulate:MARKer:FUNCtion:TOI:SEARchsignal ONCE.................................................. 800
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult?......................................................... 800
CALCulate<n>:MARKer<m>:FUNCtion:TOI[:STATe] <State>
This command initiates a measurement to determine the third intercept point.
A two-tone signal with equal carrier levels is expected at the RF input of the instru-
ment. Marker 1 and marker 2 (both normal markers) are set to the maximum of the two
signals. Delta marker 3 and delta marker 4 are positioned to the intermodulation prod-
ucts. The delta markers can be modified separately afterwards with the
CALCulate<n>:DELTamarker<m>:X command.
The third-order intercept is calculated from the level spacing between the normal mark-
ers and the delta markers.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:FUNC:TOI ON
Switches on the measurement of the third-order intercept.
Manual operation: See "TOI" on page 114
Configuring and Performing Measurements