Remote Commands
R&S
®
FSW
943User Manual 1173.9411.02 ─ 19
Usage: Query only
Manual operation: See "Marker 1/2/3/4" on page 282
See "Marker 1/2/3" on page 287
CALCulate<n>:DELTamarker<m>:Y?
This command queries the relative position of a delta marker on the y-axis.
If necessary, the command activates the delta marker first.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also INITiate:CONTinuous on page 694.
The unit depends on the application of the command.
Table 11-9: Base unit
Parameter, measuring function or result display Output unit
DBM | DBPW | DBUV | DBMV | DBUA dB (lin/log)
WATT | VOLT | AMPere dB (lin), % (log)
statistics function (APD or CCDF) on dimensionless output
Return values:
<Position> Position of the delta marker in relation to the reference marker or
the fixed reference.
Example:
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
CALC:DELT2 ON
Switches on delta marker 2.
CALC:DELT2:Y?
Outputs measurement value of delta marker 2.
Usage: Query only
CALCulate<n>:MARKer<m>:Y?
This command queries the position of a marker on the y-axis.
If necessary, the command activates the marker first.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also INITiate:CONTinuous on page 694.
Return values:
<Result> Result at the marker position.
The unit is variable and depends on the one you have currently
set.