Remote Commands
R&S
®
FSW
864User Manual 1173.9411.02 ─ 19
11.7.6 Configuring the Data Input and Output
● RF Input................................................................................................................ 864
● Using External Mixers........................................................................................... 867
● Setting up Probes..................................................................................................881
● External Generator Control................................................................................... 883
● Working with Power Sensors................................................................................ 893
● Configuring the Outputs........................................................................................ 904
11.7.6.1 RF Input
INPut:ATTenuation:PROTection:RESet............................................................................864
INPut:CONNector...........................................................................................................864
INPut:COUPling.............................................................................................................865
INPut:FILTer:HPASs[:STATe]..........................................................................................865
INPut:FILTer:YIG[:STATe].............................................................................................. 866
INPut:IMPedance...........................................................................................................866
INPut:SELect.................................................................................................................866
INPut:UPORt:STATe......................................................................................................867
INPut:UPORt[:VALue]?...................................................................................................867
INPut:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
after an overload condition occured and the protection mechanism intervened.The
error status bit (bit 3 in the STAT:QUES:POW status register) and the INPUT OVLD
message in the status bar are cleared.
(See STATus:QUEStionable:POWer[:EVENt]? on page 1046 and "STA-
Tus:QUEStionable:POWer Register" on page 639).
The command works only if the overload condition has been eliminated first.
For details on the protection mechanism see chapter 6.2.1.1, "RF Input Protection",
on page 314.
Usage:
Event
INPut:CONNector <ConnType>
Determines whether the RF input data is taken from the RF input connector or the
optional Analog Baseband I connector. This command is only available if the Analog
Baseband interface (R&S FSW-B71) is installed and active for input.
For more information on the Analog Baseband Interface (R&S FSW-B71) see the
R&S FSW I/Q Analyzer and I/Q Input User Manual.