Remote Commands
R&S
®
FSW
893User Manual 1173.9411.02 ─ 19
SOUR:EXT:FREQ:FACT:DEN 2
SOUR:EXT:FREQ:OFFS 100KHZ
//--------------Performing the calibration measurement -----------
//Perform a transmission measurement with direct connection between the generator
//and the analyzer and wait till the end
SENS:CORR:METH TRAN
SENS:CORR:COLL:ACQ THR; *WAI
//--------------Retrieving the calibration trace results -----------
//Retrieve the measured frequencies (10 Hz - 600 kHz)
TRAC:DATA:X? TRACE1
//Retrieve the measured power levels; = 0 between 10 Hz and 100 kHz (below
//generator minimum frequency); nominal -5dBm as of 100 kHz;
TRAC:DATA? TRACE1
//--------------Normalizing the calibration trace results -----------
//Retrieve the normalized power levels (= power offsets from calibration results)
//Should be 0 for all sweep points directly after calibration
SENS:CORR:STAT ON
TRAC:DATA? TRACE1
//--------------Changing the display of the calibration results -----------
//Shift the reference line so the -5 dB level is displayed in the center
DISP:TRAC:Y:SCAL:RVAL -5DB
DISP:TRAC:Y:SCAL:RPOS 50PCT
11.7.6.5 Working with Power Sensors
The following commands describe how to work with power sensors.
Configuring Power Sensors.................................................................................. 893
Configuring Power Sensor Measurements........................................................... 895
Triggering with Power Sensors............................................................................. 902
Configuring Power Sensors
SYSTem:COMMunicate:RDEVice:PMETer<p>:CONFigure:AUTO[:STATe]......................... 893
SYSTem:COMMunicate:RDEVice:PMETer:COUNt?..........................................................894
SYSTem:COMMunicate:RDEVice:PMETer<p>:DEFine......................................................894
SYSTem:COMMunicate:RDEVice:PMETer<p>:CONFigure:AUTO[:STATe] <State>
This command turns automatic assignment of a power sensor to the power sensor
index on and off.
Setting Basic Measurement Parameters