Remote Commands
R&S
®
FSW
894User Manual 1173.9411.02 ─ 19
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
SYST:COMM:RDEV:PMET:CONF:AUTO OFF
Manual operation: See "Select" on page 325
SYSTem:COMMunicate:RDEVice:PMETer:COUNt?
This command queries the number of power sensors currently connected to the
R&S FSW.
Parameters:
<NumberSensors> Number of connected power sensors.
Example:
SYST:COMM:RDEV:PMET:COUN?
Usage: Query only
Manual operation: See "Select" on page 325
SYSTem:COMMunicate:RDEVice:PMETer<p>:DEFine <Placeholder>, <Type>,
<Interface>, <SerialNo>
This command assigns the power sensor with the specified serial number to the
selected power sensor index (configuration).
The query returns the power sensor type and serial number of the sensor assigned to
the specified index.
Suffix:
<p>
.
1...4
Power sensor index
Setting parameters:
<Placeholder> Currently not evaluated
<SerialNo> Serial number of a connected power sensor
Query parameters:
<Type> The power sensor type, e.g. "NRP-Z81".
<Interface> Currently not evaluated
Return values:
<Placeholder> Currently not used
<Type> Detected power sensor type, e.g. "NRP-Z81".
<Interface> Interface the power sensor is connected to; always "USB"
Setting Basic Measurement Parameters