Remote Commands
R&S
®
FSW
865User Manual 1173.9411.02 ─ 19
Parameters:
<ConnType> RF
RF input connector
AIQI
Analog Baseband I connector
*RST: RF
Example:
INP:CONN:AIQI
Selects the analog baseband input.
Usage: SCPI confirmed
Manual operation: See "Input Connector" on page 320
INPut:COUPling <CouplingType>
This command selects the coupling type of the RF input.
Parameters:
<CouplingType> AC
AC coupling
DC
DC coupling
*RST: AC
Example:
INP:COUP DC
Usage: SCPI confirmed
Manual operation: See "Input Coupling" on page 319
INPut:FILTer:HPASs[:STATe] <State>
Activates an additional internal high-pass filter for RF input signals from 1 GHz to
3 GHz. This filter is used to remove the harmonics of the R&S FSW in order to mea-
sure the harmonics for a DUT, for example.
This function requires option R&S FSW-B13.
(Note: for RF input signals outside the specified range, the high-pass filter has no
effect. For signals with a frequency of approximately 4 GHz upwards, the harmonics
are suppressed sufficiently by the YIG filter.)
Parameters:
<State> ON | OFF
*RST: OFF
Usage: SCPI confirmed
Manual operation: See "High-Pass Filter 1...3 GHz" on page 320