Remote Commands
R&S
®
FSW
867User Manual 1173.9411.02 ─ 19
Parameters:
<Source> RF
Radio Frequency ("RF INPUT" connector)
DIQ
Digital IQ data (only available with optional Digital Baseband
Interface R&S FSW-B17)
For details on I/Q input see the R&S FSW I/Q Analyzer User
Manual.
AIQ
Analog Baseband signal (only available with optional Analog
Baseband Interface R&S FSW-B71)
For details on Analog Baseband input see the R&S FSW I/Q
Analyzer User Manual.
*RST:
RF
Manual operation: See "Radio Frequency State" on page 319
INPut:UPORt:STATe <State>
This command toggles the control lines of the user ports for the AUX PORT connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S FSW.
See the R&S FSW Getting Started manual for details.
Parameters:
<State> ON | 1
User port is switched to INPut
OFF | 0
User port is switched to OUTPut
*RST: 1
INPut:UPORt[:VALue]?
This command queries the control lines of the user ports.
Example:
INP:UPOR?
Usage: Query only
11.7.6.2 Using External Mixers
The commands required to work with external mixers in a remote environment are
described here. Note that these commands require the R&S FSW-B21 option to be
installed and an external mixer to be connected to the front panel of the R&S FSW. In
MSRA/MSRT mode, external mixers are not supported.
For details on working with external mixers see chapter 6.2.5.1, "Basics on External
Mixers", on page 358.
Setting Basic Measurement Parameters