Remote Commands
R&S
®
FSW
904User Manual 1173.9411.02 ─ 19
[SENSe:]PMETer<p>:TRIGger[:STATe] <State>
This command turns the external power trigger on and off.
This command requires the use of an R&S NRP-Z81 power sensor.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | OFF
*RST: OFF
Example:
PMET2:TRIG ON
Switches the external power trigger on
Manual operation: See "Using the power sensor as an external trigger"
on page 327
11.7.6.6
Configuring the Outputs
Configuring trigger input/output is described in chapter 11.7.4.3, "Configuring the Trig-
ger Output", on page 858.
DIAGnostic:SERVice:NSOurce........................................................................................904
OUTPut:IF[:SOURce]..................................................................................................... 904
OUTPut:IF:IFFRequency................................................................................................ 905
OUTPut:IF:SBANd?....................................................................................................... 905
OUTPut:UPORt:STATe.................................................................................................. 906
OUTPut:UPORt[:VALue].................................................................................................906
DIAGnostic:SERVice:NSOurce <State>
This command turns the 28 V supply of the BNC connector labeled NOISE SOURCE
CONTROL on the front panel on and off.
For details see chapter 6.2.1.4, "Input from Noise Sources", on page 316.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
DIAG:SERV:NSO ON
Manual operation: See "Noise Source" on page 383
OUTPut:IF[:SOURce] <Source>
Defines the type of signal available at the IF/VIDEO/DEMOD or IF OUT connector on
the rear panel of the R&S FSW.
Setting Basic Measurement Parameters