Remote Commands
R&S
®
FSW
905User Manual 1173.9411.02 ─ 19
For restrictions and more information see chapter 6.2.1.6, "IF and Video Signal Out-
put", on page 317.
Parameters:
<Source> IF
The measured IF value is available at the IF/VIDEO/DEMOD
output connector.
The frequency at which the IF value is provided is defined using
the OUTPut:IF:IFFRequency command.
IF2
The measured IF value is available at the IF OUT output con-
nector at a frequency of 2 GHz and with a bandwidth of 2 GHz.
This parameter is only available for instrument models
R&S FSW43/50/67.
VIDeo
The displayed video signal (i.e. the filtered and detected IF sig-
nal, 200mV) is available at the IF/VIDEO/DEMOD output con-
nector.
This setting is required to provide demodulated audio frequen-
cies at the output.
*RST:
IF
Example:
OUTP:IF VID
Selects the video signal for the IF/VIDEO/DEMOD output con-
nector.
Manual operation: See "IF/VIDEO/DEMOD Output" on page 382
OUTPut:IF:IFFRequency <Frequency>
This command defines the frequency for the IF output. The IF frequency of the signal is
converted accordingly.
This command is available in the time domain and if the IF/VIDEO/DEMOD output is
configured for IF.
For more information see chapter 6.2.1.6, "IF and Video Signal Output", on page 317.
Parameters:
<Frequency> *RST: 50.0 MHz
Manual operation: See "IF (Wide) Out Frequency" on page 383
OUTPut:IF:SBANd?
This command queries the sideband provided at the IF OUT connector compared to
the sideband of the RF signal. The sideband depends on the current center frequency.
This command is available only if the output is configured for IF2 (see OUTPut:IF[:
SOURce] on page 904).
For more information and prerequisites see "IF 2 GHz Output" on page 318.
Setting Basic Measurement Parameters