Remote Commands
R&S
®
FSW
903User Manual 1173.9411.02 ─ 19
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Hysteresis> Range: 3 dB to 50 dB
Increment: 1 dB
*RST: 0 dB
Example:
PMET2:TRIG:HYST 10
Sets the hysteresis of the trigger to 10 dB.
Manual operation: See "Hysteresis" on page 328
[SENSe:]PMETer<p>:TRIGger:LEVel <Level>
This command defines the trigger level for external power triggers.
This command requires the use of an R&S NRP-Z81 power sensor.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Level> -20 to +20 dBm
Range: -20 dBm to 20 dBm
*RST: -10 dBm
Example:
PMET2:TRIG:LEV -10 dBm
Sets the level of the trigger
Manual operation: See "External Trigger Level" on page 328
[SENSe:]PMETer<p>:TRIGger:SLOPe <Edge>
This command selects the trigger condition for external power triggers.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Edge> POSitive
The measurement starts in case the trigger signal shows a posi-
tive edge.
NEGative
The measurement starts in case the trigger signal shows a neg-
ative edge.
*RST: POSitive
Example:
PMET2:TRIG:SLOP NEG
Manual operation: See "Slope" on page 328