Remote Commands
R&S
®
FSW
1042User Manual 1173.9411.02 ─ 19
For details on operating modes see chapter 4, "Applications and Operating Modes",
on page 96.
Parameters:
<OpMode> SANalyzer
(Default:) Defines Signal and Spectrum Analyzer operating
mode as the presetting.
MSRA
Defines Multi-Standard Radio Analysis (MSRA) as the preset
default operating mode.
RTSM
Defines Multi-Standard Realtime (MSRT) as the preset default
operating mode.
*RST: SAN
Usage: Event
Manual operation: See "Preset Mode" on page 582
SYSTem:SECurity[:STATe] <State>
Activates or queries secure user mode.
Note: Before you activate secure user mode, store any instrument settings that are
required beyond the current session, such as predefined instrument settings, trans-
ducer files, or self-alignment data.
Note: Initially after installation of the R&S FSW-K33 option, secure user mode must be
enabled manually once before remote control is possible. This is necessary to prompt
for a change of passwords.
For details on the secure user mode see chapter 3.1.7, "Protecting Data Using the
Secure User Mode", on page 37.
Parameters:
<State> ON | OFF | 0 | 1
ON | 1
The R&S FSW automatically reboots and starts in secure user
mode. In secure user mode, no data is written to the instru-
ment's internal solid-state drive. Data that the R&S FSW nor-
mally stores on the solid-state drive is redirected to SDRAM.
OFF | 0
The R&S FSW is set to normal instrument mode. Data is stored
to the internal solid-state drive.
*RST: 0
Manual operation: See "Secure User Mode" on page 582