Measurements
R&S
®
FSW
120User Manual 1173.9411.02 ─ 19
The displayed noise level of a signal analyzer depends on its noise figure, the selected
RF attenuation, the selected reference level, the selected resolution and video band-
width and the detector.
For details see:
●
chapter 6.4.1.2, "RF Attenuation", on page 394
●
chapter 6.4.1.1, "Reference Level", on page 393
●
chapter 6.5.1.1, "Separating Signals by Selecting an Appropriate Resolution Band-
width", on page 403
●
chapter 6.5.1.2, "Smoothing the Trace Using the Video Bandwidth", on page 404
●
chapter 7.3.1.1, "Mapping Samples to Sweep Points with the Trace Detector",
on page 452
This measurement example shows the different factors influencing the S/N ratio.
Table 5-2: Signal generator settings (e.g. R&S SMU)
Frequency 128 MHz
Level -95 dBm
1.
Preset the R&S FSW.
2. Set the center frequency to 128 MHz.
3. Set the span to 100 MHz.
4.
Set the reference level to-30
dBm.
The signal is measured with the auto peak detector and is completely hidden in the
intrinsic noise of the R&S FSW.
Fig. 5-1: Sine wave signal with low S/N ratio
Basic Measurements