Measurements
R&S
®
FSW
120User Manual 1173.9411.02 ─ 19
The displayed noise level of a signal analyzer depends on its noise figure, the selected
RF attenuation, the selected reference level, the selected resolution and video band-
width and the detector.
For details see:
chapter 6.4.1.2, "RF Attenuation", on page 394
chapter 6.4.1.1, "Reference Level", on page 393
chapter 6.5.1.1, "Separating Signals by Selecting an Appropriate Resolution Band-
width", on page 403
chapter 6.5.1.2, "Smoothing the Trace Using the Video Bandwidth", on page 404
chapter 7.3.1.1, "Mapping Samples to Sweep Points with the Trace Detector",
on page 452
This measurement example shows the different factors influencing the S/N ratio.
Table 5-2: Signal generator settings (e.g. R&S SMU)
Frequency 128 MHz
Level -95 dBm
1.
Preset the R&S FSW.
2. Set the center frequency to 128 MHz.
3. Set the span to 100 MHz.
4.
Set the reference level to-30
dBm.
The signal is measured with the auto peak detector and is completely hidden in the
intrinsic noise of the R&S FSW.
Fig. 5-1: Sine wave signal with low S/N ratio
Basic Measurements