EasyManua.ls Logo

Rohde & Schwarz FSW43 - Page 250

Rohde & Schwarz FSW43
1150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurements
R&S
®
FSW
250User Manual 1173.9411.02 ─ 19
File contents Explanation
Version;1.00; Firmware version
Date;31.Mar 11; Storage date of data set
Mode;ANALYZER; SPURIOUS; Operating mode and measurement function
Center Freq;13250000000.000000;Hz X-axis settings
Freq Offset;0.000000;Hz
Span;26499982000.000000;Hz
x-Axis;LIN;
Start;9000.000000;Hz
Stop;8000000000.000000;Hz
Level Offset;0.000000;dB Y-axis settings
Ref Position;100.000000;%
y-Axis;LOG;
Level Range;100.000000;dB
Trace settings
Trace Mode;CLR/WRITE;
Sweep Count;1;
TRACE 1:
Trace Mode;CLR/WRITE;
x-Unit;Hz;
y-Unit;dBm;
List evaluation settings
Margin;6.000000;s Peak List margin
PeaksPerRange;25; Max. number of peaks per range to be detected
Values;3; Number of detected peaks
File data section
0;9000;150000;1000;79500;-25.006643295288086;-
12.006643295288086;PASS;
0;9000;150000;1000;101022.11126961483;-47.075
111389160156;-34.075111389160156;PASS;
0;9000;150000;1000;58380.171184022824;-47.079
341888427734;-34.079341888427734;PASS;
Measured peak values:
<range number>;
<start frequency>;
<stop frequency>;
<resolution bandwidth of range>;
<frequency of peak>;
<absolute power in dBm of peak>;
<distance to the limit line in dB>; (positive value
means above the limit)
<limit fail (pass = 0, fail =1)>;
Spurious Emissions Measurement

Table of Contents