Common Analysis and Display Functions
R&S
®
FSW
468User Manual 1173.9411.02 ─ 19
Function Trace Settings
Preset All Traces Trace 1: Clear Write
Auto Detector (Auto Peak)
Traces 2-6: Blank
Auto Detector
Set Trace Mode
Max | Avg | Min
Trace 1: Max Hold
Auto Detector (Positive Peak)
Trace 2: Average
Auto Detector (Sample)
Trace 3: Min Hold
Auto Detector (Negative Peak)
Traces 4-6: Blank
Auto Detector
Set Trace Mode
Max | ClrWrite | Min
Trace 1: Max Hold
Auto Detector (Positive Peak)
Trace 2: Clear Write
Auto Detector (Auto Peak)
Trace 3: Min Hold
Auto Detector (Negative Peak)
Traces 4-6: Blank
Auto Detector
Trace 1/Trace 2/Trace 3/Trace 4 (Softkeys)
Displays the "Traces" settings and focuses the "Mode" list for the selected trace.
For details see chapter 7.3.3.1, "How to Configure a Standard Trace", on page 477.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>[:STATe] on page 911
Copy Trace
The "Copy Trace" softkey opens the "Copy Trace" tab of the "Trace Configuration" dia-
log box.
The "Copy Trace" tab contains functionality to copy trace data to another trace.
The first group of buttons (labelled "Trace 1" to "Trace 6") select the source trace. The
second group of buttons (labelled "Copy to Trace 1" to "Copy to Trace 6") select the
destination.
Remote command:
TRACe<n>:COPY on page 913