General Instrument Setup
R&S
®
FSW
575User Manual 1173.9411.02 ─ 19
Comment
Defines an optional comment for the transducer line. The text may contain up to 40
characters.
Remote command:
[SENSe:]CORRection:TRANsducer:COMMent on page 1026
Unit
The unit in which the y-values of the data points of the transducer line are defined.
As soon as a transducer is activated, the unit of the transducer is automatically used
for all the level settings and outputs. The unit cannot be changed in the amplitude set-
tings unless dB is used.
Remote command:
[SENSe:]CORRection:TRANsducer:UNIT on page 1027
X-Axis Scaling
Describes the scaling of the horizontal axis on which the data points of the transducer
line are defined. Scaling can be linear or logarithmic.
Remote command:
[SENSe:]CORRection:TRANsducer:SCALing on page 1026
Data points
Each transducer line is defined by a minimum of 2 and a maximum of 50 data points.
Each data point is defined by its position (x-axis) and value (y-value).
The data points must comply with the following rules to ensure correct operation:
The frequencies for the data points must always be defined in ascending order.
Otherwise the entry will not be accepted and the an error message is displayed.
The frequencies of the data points may exceed the valid frequency range of the
R&S FSW since only the set frequency range is taken into account for measure-
ments. The minimum frequency of a data point is 0 Hz, the maximum frequency
200 GHz.
The value range for the transducer factor is ±200 dB.
Gain has to be entered as a negative value, and attenuation as a positive value.
Remote command:
[SENSe:]CORRection:TRANsducer:DATA on page 1026
Insert Value
Inserts a data point in the transducer line above the selected one in the "Edit Trans-
ducer" dialog box.
Delete Value
Deletes the selected data point in the "Edit Transducer" dialog box.
Shift x
Shifts the x-value of each data point horizontally by the defined shift width.
Shift y
Shifts the y-value of each data point vertically by the defined shift width.
General Instrument Settings