EasyManua.ls Logo

Rohde & Schwarz FSW43 - Page 745

Rohde & Schwarz FSW43
1150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSW
745User Manual 1173.9411.02 ─ 19
Parameters:
<State> ABSolute | RELative | AND | OR
ABSolute
Checks only the absolute limits defined.
RELative
Checks only the relative limits. Relative limits are defined as rel-
ative to the measured power in the reference range.
AND
Combines the absolute and relative limit. The limit check fails
when both limits are violated.
OR
Combines the absolute and relative limit. The limit check fails
when one of the limits is violated.
*RST:
RELative
Manual operation: See "Limit Check <n>" on page 213
[SENSe:]ESPectrum<sb>:RANGe<range>:POINts:MIN[:VALue] <SweepPoint>
Defines the minimum number of sweep points for the range.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1..30
Selects the measurement range.
Parameters:
<SweepPoints> Minimum number of sweep points per range
Range: 1 to 32001
*RST: 1
Example:
SENSe1:ESPectrum:RANGe3:POINts:MIN:VALue 400
Manual operation: See "Minimum Sweep Points" on page 214
[SENSe:]ESPectrum<sb>:RANGe<range>:MLCalc <Function>
Defines the function used to calculate the limit line for the n-th power class for overlap-
ping ranges in Multi-SEM measurements. For details see "Limit calculation for individ-
ual ranges" on page 206.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1..30
Selects the measurement range.