Remote Commands
R&S
®
FSW
752User Manual 1173.9411.02 ─ 19
Parameters:
<NoPowerClasses> 1 to 4
*RST: 1
Example:
CALC:LIM:ESP:PCL:COUN 2
Two power classes can be defined.
Manual operation: See "Adding or Removing a Power Class" on page 219
CALCulate:LIMit:ESPectrum<sb>:PCLass<class>[:EXCLusive] <State>
This command selects the power class used by the measurement if CALCulate:
LIMit:ESPectrum<sb>:MODE is set to manual.
Note that:
●
You can only use power classes for which limits are defined.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<class> 1...4
power class
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:LIM:ESP:PCL1 ON
Activates the first defined power class.
Manual operation: See "Used Power Classes" on page 218
CALCulate:LIMit:ESPectrum<sb>:PCLass<class>:LIMit[:STATe] <State>
This command selects the limit check mode for each power class.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<class> 1...4
power class
Configuring and Performing Measurements